Journal of Heart Valve Disease 2003-Jan

Kinematics of synthetic flexible leaflet heart valves during accelerated testing.

Само регистровани корисници могу преводити чланке
Пријави се / Пријави се
Веза се чува у привремену меморију
Stephen S D'Souza
Maya Butterfield
John Fisher

Кључне речи

Апстрактан

OBJECTIVE

The study aim was to examine the effect of variation of test frequencies on the duration and kinematics of valve leaflets during opening, and at the fully open position.

METHODS

Trileaflet polyurethane heart valves were tested in a Rowan Ash fatigue accelerator at 5, 10 and 15 Hz. Images of the valves were acquired using a high-speed camera and assessed qualitatively during opening and in the fully open position. The bending strains along the leaflet free edge and leaflet belly were also calculated from these images. Results were compared with those found at 1 Hz in a pulsatile flow simulator. The time duration of each phase of the cycle and the strain time integral when opening were also determined.

RESULTS

In the fully open position, the maximum bending strain found at the commissural regions of the valve was not dependent on the cycle rate. During opening, one leaflet always deformed more than either of the others, but this leaflet was not necessarily the same one each time. The opening characteristics of the valves were dependent on cycle rate: increasing bending was observed in the belly of the leaflet with increasing frequency. The ratio of time of opening of the leaflets to time whilst fully open increased with the cycle rate, and the period of time when the valve was fully open and fully closed reduced markedly with increased cycle rate.

CONCLUSIONS

Leaflet kinematics, bending strains, the proportion of time the valve was fully closed or open, and the strain time integral were dependent on cycle rate in accelerated testing. This may influence the relative severity of accelerated testers.

Придружите се нашој
facebook страници

Најкомплетнија база лековитог биља подржана науком

  • Ради на 55 језика
  • Биљни лекови потпомогнути науком
  • Препознавање биљака по слици
  • Интерактивна ГПС мапа - означите биље на локацији (ускоро)
  • Читајте научне публикације повезане са вашом претрагом
  • Претражите лековито биље по њиховим ефектима
  • Организујте своја интересовања и будите у току са истраживањем вести, клиничким испитивањима и патентима

Упишите симптом или болест и прочитајте о биљкама које би могле да помогну, укуцајте неку биљку и погледајте болести и симптоме против којих се користи.
* Све информације се заснивају на објављеним научним истраживањима

Google Play badgeApp Store badge