Journal of Clinical and Diagnostic Research 2014-Jul

Comparison of Point of Care (POC) Testing of Glucose by B Braun Glucometer and Hemocue Glucose 201+ Analyser Versus Centralised Testing in Neonatal Intensive Care Unit (NICU).

Endast registrerade användare kan översätta artiklar
Logga in Bli medlem
Länken sparas på Urklipp
Sudha Reddy V R
Sumathi M E
Beere Gowda Y C
Mohamed Suhail S

Nyckelord

Abstrakt

BACKGROUND

Neonatal hypoglycemia is the most common carbohydrate metabolic disturbance seen in case of neonates and especially in preterm neonates. Accurate and rapid determination of hypoglycemia and its prompt treatment is of utmost importance to decrease morbidity and mortality of neonates.

OBJECTIVE

To estimate blood glucose in neonates and test the efficacy of HemoCue Glucose 201+ analyser and B Braun Glucometer by comparing with centralised laboratory testing. To compare the blood glucose in capillary and venous blood samples of neonates.

METHODS

Hospital setting; Comparative Study Materials and Methods: After obtaining informed consent, all neonates admitted to Neonatal Intensive Care Unit (NICU) were screened for blood sugar. Capillary and venous blood glucose was estimated employing HemoCue Glucose 201+ analyser and B Braun Glucometer. Simultaneously, the same venous sample was collected in fluoride tube and sent to central clinical biochemistry laboratory for glucose estimation. When anaemia or polycythemia was clinically suspected the same venous sample was sent for estimation of Hematocrit (Hct).

METHODS

Comparison of blood glucose concentration of B Braun glucometer, HemoCue Glucose 201+ analyser and centralis/ed plasma glucose levels was done by using students test. All the statistical analysis were done using software SPSS 6 version.

RESULTS

Mean values of blood glucose (100.2 + 48.4) with B Braun glucometer was significantly higher (p=0.003) when compared to plasma glucose values (76.95 + 45.99) estimated in central laboratory and HemoCue glucose 201+ analyser (82.9 + 51.4). HemoCue glucose 201+ analyser did not show significant difference (p=0.463) with central laboratory testing. There was no significant difference between the capillary and venous sample estimated in both the instruments. Estimation with HemoCue glucose 201+ analyser correlated well with central laboratory testing in neonates with blood glucose <55mg/dl,

CONCLUSIONS

We conclude that HemoCue glucose 201+ analyser appears to be a suitable point of care (POC) blood glucose measurement device in neonates on both capillary and venous blood samples, as it showed a good correlation with central laboratory values without significant interference from Hct.

Gå med på vår
facebook-sida

Den mest kompletta databasen med medicinska örter som stöds av vetenskapen

  • Fungerar på 55 språk
  • Växtbaserade botemedel som stöds av vetenskap
  • Örter igenkänning av bild
  • Interaktiv GPS-karta - märka örter på plats (kommer snart)
  • Läs vetenskapliga publikationer relaterade till din sökning
  • Sök efter medicinska örter efter deras effekter
  • Organisera dina intressen och håll dig uppdaterad med nyheterna, kliniska prövningar och patent

Skriv ett symptom eller en sjukdom och läs om örter som kan hjälpa, skriv en ört och se sjukdomar och symtom den används mot.
* All information baseras på publicerad vetenskaplig forskning

Google Play badgeApp Store badge