Journal of Invasive Cardiology 2017-Jan

Initial Single-Center Experience With the Fully Repositionable Transfemoral Lotus Aortic Valve System.

Endast registrerade användare kan översätta artiklar
Logga in Bli medlem
Länken sparas på Urklipp
Kai-Uwe Jarr
Florian Leuschner
Benjamin Meder
Hugo A Katus
Raffi Bekeredjian
Emmanuel Chorianopoulos

Nyckelord

Abstrakt

BACKGROUND

Transcatheter aortic valve replacement (TAVR) has become the standard therapy for patients with severe symptomatic aortic stenosis and unacceptable high risk for surgical aortic valve replacement. Several different devices for TAVR have been introduced so far, each of them with unique features.

OBJECTIVE

To analyze our first real-world experience with the second- generation, mechanically expanded, fully repositionable transfemoral Lotus TAVR device (Boston Scientific).

RESULTS

Between August 2014 and January 2016, a total of 298 patients were scheduled for transfemoral TAVR at our institution. Among them, 145 patients received a CoreValve/CoreValve Evolut R device (Medtronic), 90 patients received a Sapien 3 device (Edwards Lifesciences), and 63 patients were scheduled for Lotus TAVR device implantation based on access vessel size and annulus diameter. Of the 63 subsequent patients scheduled, 62 were successfully implanted. Short-term Valve Academic Research Consortium (VARC)-2 results at 30 days showed a 6.4% all-cause mortality, a 3.2% major stroke rate, a 36.5% new permanent pacemaker rate and 1.6% major vascular complications. Of note was the near absence of significant residual aortic regurgitation in the presence of a favorable hemodynamic profile even in patients with difficult anatomy.

CONCLUSIONS

Our first real-time experience with the fully repositionable Lotus TAVR system mimics the results of early clinical studies like REPRISE II. Further next-generation modifications in combination with gaining experience with this new device will likely reduce the need for new permanent pacemaker implantation and vascular complications in the future.

Gå med på vår
facebook-sida

Den mest kompletta databasen med medicinska örter som stöds av vetenskapen

  • Fungerar på 55 språk
  • Växtbaserade botemedel som stöds av vetenskap
  • Örter igenkänning av bild
  • Interaktiv GPS-karta - märka örter på plats (kommer snart)
  • Läs vetenskapliga publikationer relaterade till din sökning
  • Sök efter medicinska örter efter deras effekter
  • Organisera dina intressen och håll dig uppdaterad med nyheterna, kliniska prövningar och patent

Skriv ett symptom eller en sjukdom och läs om örter som kan hjälpa, skriv en ört och se sjukdomar och symtom den används mot.
* All information baseras på publicerad vetenskaplig forskning

Google Play badgeApp Store badge