Screening for amblyopia in preverbal children with photoscreening photographs. III. improved grading criteria for hyperopia.

Endast registrerade användare kan översätta artiklar
Logga in Bli medlem
Länken sparas på Urklipp
P Y Tong
J P Macke
R E Bassin
M Everett
E Enke-Miyazaki
J M Tielsch
D R Stager
M M Parks
G R Beauchamp

Nyckelord

Abstrakt

OBJECTIVE

To examine the ability of the Medical Technology and Innovations (MTI), Inc., Photoscreener (Cedar Falls, IA) to detect hyperopia and to improve the photograph grading criteria to screen for amblyopiogenic levels of hyperopia.

METHODS

Cross-sectional study and reanalysis. PARTICIPANTS AND TESTING: In previous work, 392 participants received a complete ophthalmologic examination and were photographed using the MTI Photoscreener. For this study, all 209 participants with normal examination findings (65 children) or hyperopia without anisometropia (144 children) were selected. The data were reanalyzed using modified photograph grading and ophthalmologic examination failure criteria. Potential reasons for why many children with hyperopia passed photoscreening were explored.

METHODS

We determined whether a study participant would pass or fail screening with a given photograph grading and ophthalmologic examination failure criteria.

RESULTS

Most children with hyperopia of +2.00 to +3.50 diopters (D) passed screening with the MTI instrument, in most cases because their photographs lacked bright crescents. When bright crescents in at least two of the four possible meridians were the grading guideline for screening failure and the pediatric ophthalmologists' consensus hyperopia failure criteria (> +3.50 D) were adopted, the sensitivity for hyperopia detection was 100% and the specificity was 88%. Identical results were obtained using the American Academy of Ophthalmology Preferred Practice Pattern hyperopia failure criteria (>/= +4.50 D).

CONCLUSIONS

The MTI photograph grading guidelines can be simplified, and the ophthalmologic examination failure criteria for hyperopia can be improved. The presence of a bright crescent in the lower or the left pupillary margin indicate hyperopia in an amblyopiogenic range (> +3.50 D).

Gå med på vår
facebook-sida

Den mest kompletta databasen med medicinska örter som stöds av vetenskapen

  • Fungerar på 55 språk
  • Växtbaserade botemedel som stöds av vetenskap
  • Örter igenkänning av bild
  • Interaktiv GPS-karta - märka örter på plats (kommer snart)
  • Läs vetenskapliga publikationer relaterade till din sökning
  • Sök efter medicinska örter efter deras effekter
  • Organisera dina intressen och håll dig uppdaterad med nyheterna, kliniska prövningar och patent

Skriv ett symptom eller en sjukdom och läs om örter som kan hjälpa, skriv en ört och se sjukdomar och symtom den används mot.
* All information baseras på publicerad vetenskaplig forskning

Google Play badgeApp Store badge