International Journal of Hyperthermia 2004-Aug

A feasibility study of interstitial hyperthermia plus external beam radiotherapy in glioblastoma multiforme using the Multi ELectrode Current Source (MECS) system.

Chỉ người dùng đã đăng ký mới có thể dịch các bài báo
Đăng nhập Đăng ký
Liên kết được lưu vào khay nhớ tạm
M C C M Hulshof
B W Raaymakers
J J W Lagendijk
R W Koot
H Crezee
L J A Stalpers
D González González

Từ khóa

trừu tượng

OBJECTIVE

Thermoradiotherapy has been shown in several randomized trials to increase local control compared to radiotherapy alone. The first randomized study of interstitial hyperthermia in glioblastoma multiforme showed a survival benefit for hyperthermia, though small. Improvement of the heating technique could lead to improved results. The purpose of this feasibility study is to present the clinical and thermal data of application of an improved interstitial hyperthermia system.

METHODS

Six patients with a glioblastoma multiforme were treated with interstitial hyperthermia using the Multi Electrode Current Source Interstitial Hyperthermia (MECS-IHT) system. The MECS-IHT system has the capability of spatial monitoring of temperature and individually steering of heating electrodes. Three sessions were given aiming at a steady state temperature of 42 degrees C for 1 h, with an interval of 3-4 days, during an external irradiation scheme of 60 Gy in 6 weeks. Hyperthermia was delivered with a mean of 10 catheters, 18 heating electrodes and 38 thermal probes per patient.

RESULTS

Sub-optimal temperatures were encountered in the first two patients leading to adjustments in technique thereafter with subsequent improvement of thermal data. With a catheter spacing of 11-12 mm, measurements yielded a mean T(90), T(50) and T(10) of 39.9, 43.7 and 45.2 degrees C, respectively, over three sessions in the last patient. The power per electrode to reach this temperature distribution varied from 25-100% of full power in each of the last four patients. Thermal data were reproducible over the three sessions. Acute toxicity was minimal.

CONCLUSIONS

Despite the spatial steering capabilities of the MECS-IHT system, a large temperature heterogeneity was encountered. The heterogeneity was the reason to limit the catheter spacing to 11-12 mm, thus making only small tumour volumes feasible for interstitial heating.

Tham gia trang
facebook của chúng tôi

Cơ sở dữ liệu đầy đủ nhất về dược liệu được hỗ trợ bởi khoa học

  • Hoạt động bằng 55 ngôn ngữ
  • Phương pháp chữa bệnh bằng thảo dược được hỗ trợ bởi khoa học
  • Nhận dạng các loại thảo mộc bằng hình ảnh
  • Bản đồ GPS tương tác - gắn thẻ các loại thảo mộc vào vị trí (sắp ra mắt)
  • Đọc các ấn phẩm khoa học liên quan đến tìm kiếm của bạn
  • Tìm kiếm dược liệu theo tác dụng của chúng
  • Sắp xếp sở thích của bạn và cập nhật các nghiên cứu tin tức, thử nghiệm lâm sàng và bằng sáng chế

Nhập một triệu chứng hoặc một căn bệnh và đọc về các loại thảo mộc có thể hữu ích, nhập một loại thảo mộc và xem các bệnh và triệu chứng mà nó được sử dụng để chống lại.
* Tất cả thông tin dựa trên nghiên cứu khoa học đã được công bố

Google Play badgeApp Store badge