Avecor Trillium oxygenator versus noncoated Monolyth oxygenator: a prospective randomized controlled study.

Chỉ người dùng đã đăng ký mới có thể dịch các bài báo
Đăng nhập Đăng ký
Liên kết được lưu vào khay nhớ tạm
Frédéric Vanden Eynden
Michel Carrier
Sandra Ouellet
Philippe Demers
Jessica Forcillo
Louis P Perrault
Michel Pellerin
Denis Bouchard

Từ khóa

trừu tượng

OBJECTIVE

The surface coating of a synthetic surface is currently investigated to decrease the harmful effects of cardiopulmonary bypass (CPB). This study was designed to study the effects of the surface coating of a hollow fiber membrane oxygenator on coagulation, inflammation markers, and clinical outcomes. The biomaterials used to coat the membrane include heparin, polyethylene oxide chains (PEO), and sulfate/sulfonate groups. The coated membrane was compared to an uncoated oxygenator made of polypropylene.

METHODS

Two hundred patients who were scheduled to undergo valve repair and/or replacement surgery with or without coronary surgery were enrolled in the study. The patients were randomized to undergo CPB with either the Avecor oxygenator with Trillium (Medtronic, Minneapolis, MN, USA), a biopassive surface, or the Monolyth (Sorin, Irvine, CA, USA) oxygenator without coating. The primary and secondary endpoints were the differences between these oxygenators in regard to patients' biochemistry, coagulation profiles, inflammatory mediators, and clinical outcomes, including blood loss and neurological events.

RESULTS

There were no differences between the two groups in terms of biochemistry, coagulation profile, inflammatory mediator release, and blood loss. Five patients in the Avecor group showed clinical evidence of a stroke confirmed with computerized tomography (CT) scan imaging, and none in the noncoated oxygenator group.

CONCLUSIONS

The oxygenator Avecor offers similar results in terms of inflammation and coagulation profiles and blood loss during valvular surgery compared to a standard uncoated control oxygenator. The rate of neurological events was unusually elevated in the former group of patients, with only speculative explanation at this point. Further studies are warranted to clarify this aspect.

Tham gia trang
facebook của chúng tôi

Cơ sở dữ liệu đầy đủ nhất về dược liệu được hỗ trợ bởi khoa học

  • Hoạt động bằng 55 ngôn ngữ
  • Phương pháp chữa bệnh bằng thảo dược được hỗ trợ bởi khoa học
  • Nhận dạng các loại thảo mộc bằng hình ảnh
  • Bản đồ GPS tương tác - gắn thẻ các loại thảo mộc vào vị trí (sắp ra mắt)
  • Đọc các ấn phẩm khoa học liên quan đến tìm kiếm của bạn
  • Tìm kiếm dược liệu theo tác dụng của chúng
  • Sắp xếp sở thích của bạn và cập nhật các nghiên cứu tin tức, thử nghiệm lâm sàng và bằng sáng chế

Nhập một triệu chứng hoặc một căn bệnh và đọc về các loại thảo mộc có thể hữu ích, nhập một loại thảo mộc và xem các bệnh và triệu chứng mà nó được sử dụng để chống lại.
* Tất cả thông tin dựa trên nghiên cứu khoa học đã được công bố

Google Play badgeApp Store badge