Ophthalmologe 2010-Sep

[Clinical results of the aspheric intraocular lens FY-60AD (Hoya) with particular respect to decentration and tilt].

Chỉ người dùng đã đăng ký mới có thể dịch các bài báo
Đăng nhập Đăng ký
Liên kết được lưu vào khay nhớ tạm
U Mester
S Heinen
H Kaymak

Từ khóa

trừu tượng

BACKGROUND

Aspheric intraocular lenses (IOLs) aim to improve visual function and particularly contrast vision by neutralizing spherical aberration. One drawback of such IOLs is the enhanced sensitivity to decentration and tilt, which can deteriorate image quality.

METHODS

A total of 30 patients who received bilateral phacoemulsification before implantation of the aspheric lens FY-60AD (Hoya) were included in a prospective study. In 25 of the patients (50 eyes) the following parameters could be assessed 3 months after surgery: visual acuity, refraction, contrast sensitivity, pupil size, wavefront errors and decentration and tilt using a newly developed device.

RESULTS

The functional results were very satisfying and comparable to results gained with other aspheric IOLs. The mean refraction was sph + 0.1 D (±0.7 D) and cyl 0.6 D (±0.8 D). The spherical equivalent was −0.2 D (±0.6 D). Wavefront measurements revealed a good compensation of the corneal spherical aberration but vertical and horizontal coma also showed opposing values in the cornea and IOL. The assessment of the lens position using the Purkinje meter demonstrated uncritical amounts of decentration and tilt. The mean amount of decentration was 0.2 mm±0.2 mm in the horizontal and vertical directions. The mean amount of tilt was 4.0±2.1° in horizontal and 3.0±2.5° in vertical directions.

CONCLUSIONS

In a normal dioptric power range the aspheric IOL FY-60AD compensates the corneal spherical aberration very well with only minimal decentration. The slight tilt is symmetrical in both eyes and corresponds to the position of the crystalline lens in young eyes. This may contribute to our findings of compensated corneal coma.

Tham gia trang
facebook của chúng tôi

Cơ sở dữ liệu đầy đủ nhất về dược liệu được hỗ trợ bởi khoa học

  • Hoạt động bằng 55 ngôn ngữ
  • Phương pháp chữa bệnh bằng thảo dược được hỗ trợ bởi khoa học
  • Nhận dạng các loại thảo mộc bằng hình ảnh
  • Bản đồ GPS tương tác - gắn thẻ các loại thảo mộc vào vị trí (sắp ra mắt)
  • Đọc các ấn phẩm khoa học liên quan đến tìm kiếm của bạn
  • Tìm kiếm dược liệu theo tác dụng của chúng
  • Sắp xếp sở thích của bạn và cập nhật các nghiên cứu tin tức, thử nghiệm lâm sàng và bằng sáng chế

Nhập một triệu chứng hoặc một căn bệnh và đọc về các loại thảo mộc có thể hữu ích, nhập một loại thảo mộc và xem các bệnh và triệu chứng mà nó được sử dụng để chống lại.
* Tất cả thông tin dựa trên nghiên cứu khoa học đã được công bố

Google Play badgeApp Store badge