Neonatology 2017

Reliability of Pulse Oximetry during Progressive Hypoxia, Cardiopulmonary Resuscitation, and Recovery in a Piglet Model of Neonatal Hypoxic Cardiac Arrest.

Chỉ người dùng đã đăng ký mới có thể dịch các bài báo
Đăng nhập Đăng ký
Liên kết được lưu vào khay nhớ tạm
Mohammad Ahmad Hassan
Claudia Weber
Markus Waitz
Li Huang
Helmut D Hummler
Marc Robin Mendler

Từ khóa

trừu tượng

BACKGROUND

Pulse oximetry is widely used in intensive care and emergency conditions to monitor arterial oxygenation and to guide oxygen therapy.

OBJECTIVE

To study the reliability of pulse oximetry in comparison with CO-oximetry in newborn piglets during progressive hypoxia, cardiac arrest, cardiopulmonary resuscitation (CPR), and after return of spontaneous circulation (ROSC).

METHODS

Thirty-three newborn piglets were exposed to hypoxia until asystole occurred and then resuscitated until ROSC. Arterial oxygen saturation was monitored continuously by pulse oximetry (SpO2) with one sensor applied to the wrist of the right forelimb (FL) and another to the thigh of the left hind limb (HL). Arterial functional oxygen saturation (SaO2) was measured at baseline and at predefined intervals during each phase of the experiment. SpO2 was compared with coinciding SaO2 values and bias considered whenever the difference (SpO2 - SaO2) was beyond ±5%.

RESULTS

Bias values were lower at the baseline measurements (-3.7 ± 2.3% in FL and -4.1 ± 3.4% in HL) as well as after ROSC (1.5 ± 4.2% in FL and 0.2 ± 4.6% in HL) with higher precision and accuracy than during other experiment phases. During hypoxia induction, cardiac arrest, and CPR, there was a marked decrease in precision and accuracy as well as an increase in bias up to 43 ± 26 and 56 ± 27% in FL and HL, respectively, over a range of SaO2 from 13 to 51%.

CONCLUSIONS

Pulse oximetry showed increased bias and decreased accuracy and precision during marked hypoxemia in a model of neonatal hypoxic cardiac arrest.

Tham gia trang
facebook của chúng tôi

Cơ sở dữ liệu đầy đủ nhất về dược liệu được hỗ trợ bởi khoa học

  • Hoạt động bằng 55 ngôn ngữ
  • Phương pháp chữa bệnh bằng thảo dược được hỗ trợ bởi khoa học
  • Nhận dạng các loại thảo mộc bằng hình ảnh
  • Bản đồ GPS tương tác - gắn thẻ các loại thảo mộc vào vị trí (sắp ra mắt)
  • Đọc các ấn phẩm khoa học liên quan đến tìm kiếm của bạn
  • Tìm kiếm dược liệu theo tác dụng của chúng
  • Sắp xếp sở thích của bạn và cập nhật các nghiên cứu tin tức, thử nghiệm lâm sàng và bằng sáng chế

Nhập một triệu chứng hoặc một căn bệnh và đọc về các loại thảo mộc có thể hữu ích, nhập một loại thảo mộc và xem các bệnh và triệu chứng mà nó được sử dụng để chống lại.
* Tất cả thông tin dựa trên nghiên cứu khoa học đã được công bố

Google Play badgeApp Store badge