European Journal of Radiology 2012-Dec

Triple-rule-out dual-source CT angiography of patients with acute chest pain: dose reduction potential of 100 kV scanning.

Chỉ người dùng đã đăng ký mới có thể dịch các bài báo
Đăng nhập Đăng ký
Liên kết được lưu vào khay nhớ tạm
Radko Krissak
Thomas Henzler
Anne Prechel
Miriam Reichert
Joachim Gruettner
Tim Sueselbeck
Stefan O Schoenberg
Christian Fink

Từ khóa

trừu tượng

OBJECTIVE

To investigate the dose reduction potential of low kV triple-rule-out dual-source CT angiography (TRO-CTA) in non-obese (BMI ≤ 25 kg/m(2)) patients with acute chest pain.

METHODS

Sixty consecutive patients were randomly assigned to two different retrospectively ECG-gated TRO-CTA protocols in this prospective trial: Thirty patients were examined with a 120-kV standard protocol (320 reference mAs with automatic tube current modulation, automatically adapted pitch and ECG-pulsing) and served as the control group (group 1), an otherwise identical 100 kV protocol was used in the other thirty patients (group 2) for a radiation dose reduction. Subjective image quality was assessed on a 5 point scale (1: excellent, 5: non-diagnostic) by two blinded observers. Quantitative image analysis assessed vascular attenuation, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) in different vascular segments. The effective dose was calculated from the dose length product (DLP) using a conversion coefficient of 0.017 mSvmGy(-1)cm(-1).

RESULTS

There was no significant difference of age, BMI, heart rate, pitch or scan length between both patient groups. Subjective image quality was rated similar in both groups (group 1: 1.2 ± 0.4, group 2: average score=1.3 ± 0.5). Vessel attenuation was significantly higher in group 2 than in group 1 (ascending aorta: 456 ± 83 HU vs. 370 ± 78 HU, p<0.001; pulmonary artery: 468 ± 118 HU vs. 411 ± 91 HU, p=0.03; left coronary artery: 437 ± 110 HU vs. 348 ± 89 HU, p<0.001), however, there was no significant difference in SNR (13.2 ± 7.6 vs. 14.5 ± 7.5, p=0.49) or CNR (13.8 ± 6.6 vs. 15.9 ± 7.7, p=0.25). The effective radiation dose of the 100 kV protocol was significantly lower (9.6 ± 3.2 mSv vs. 18.1 ± 9.4 mSv, p<0.0001).

CONCLUSIONS

TRO-CTA with 100 kV is feasible in non-obese patients and results in diagnostic image quality and significantly reduced radiation dose.

Tham gia trang
facebook của chúng tôi

Cơ sở dữ liệu đầy đủ nhất về dược liệu được hỗ trợ bởi khoa học

  • Hoạt động bằng 55 ngôn ngữ
  • Phương pháp chữa bệnh bằng thảo dược được hỗ trợ bởi khoa học
  • Nhận dạng các loại thảo mộc bằng hình ảnh
  • Bản đồ GPS tương tác - gắn thẻ các loại thảo mộc vào vị trí (sắp ra mắt)
  • Đọc các ấn phẩm khoa học liên quan đến tìm kiếm của bạn
  • Tìm kiếm dược liệu theo tác dụng của chúng
  • Sắp xếp sở thích của bạn và cập nhật các nghiên cứu tin tức, thử nghiệm lâm sàng và bằng sáng chế

Nhập một triệu chứng hoặc một căn bệnh và đọc về các loại thảo mộc có thể hữu ích, nhập một loại thảo mộc và xem các bệnh và triệu chứng mà nó được sử dụng để chống lại.
* Tất cả thông tin dựa trên nghiên cứu khoa học đã được công bố

Google Play badgeApp Store badge