Azerbaijani
Albanian
Arabic
Armenian
Azerbaijani
Belarusian
Bengali
Bosnian
Catalan
Czech
Danish
Deutsch
Dutch
English
Estonian
Finnish
Français
Greek
Haitian Creole
Hebrew
Hindi
Hungarian
Icelandic
Indonesian
Irish
Italian
Japanese
Korean
Latvian
Lithuanian
Macedonian
Mongolian
Norwegian
Persian
Polish
Portuguese
Romanian
Russian
Serbian
Slovak
Slovenian
Spanish
Swahili
Swedish
Turkish
Ukrainian
Vietnamese
Български
中文(简体)
中文(繁體)
Journal of Cataract and Refractive Surgery 2005-Apr

Thermal comparison of the AdvanTec Legacy, Sovereign WhiteStar, and Millennium phacoemulsification systems.

Yalnız qeydiyyatdan keçmiş istifadəçilər məqalələri tərcümə edə bilərlər
Giriş / Qeydiyyatdan keçin
Bağlantı panoya saxlanılır
Richard J Mackool
Marc A Sirota

Açar sözlər

Mücərrəd

OBJECTIVE

To determine the operating temperature of ultrasonic tips used with the Alcon AdvanTec Legacy, AMO Sovereign WhiteStar, and Bausch & Lomb Millennium phacoemulsification systems.

METHODS

Mackool Eye Institute, Astoria, New York, New York, USA.

METHODS

Thermal imaging of ultrasonic tips (Legacy, WhiteStar, and Millennium) was performed in air and in human cadaver eyes using a duty cycle of 33% (WhiteStar) and 50% (Millennium and Legacy). In vitro temperatures were measured with the tip centered in the incision and intentionally decentered against the side of the incision. The stroke length of each instrument was also measured, and the operating frequency of the Legacy was evaluated with the addition of a tangential load.

RESULTS

Open air and in vitro testing demonstrated that tip temperatures with the Legacy were consistently the lowest. Temperatures measured with the WhiteStar and Millennium systems were higher and generally similar to each other. At identical console power settings, the stroke length of the WhiteStar and Millennium tips was longer than that of the Legacy. The frequency of the Legacy handpiece did not change significantly (less than 200 Hz) under conditions of tangential tip loading.

CONCLUSIONS

At identical console power settings and similar console duty cycles, the temperature elevation of ultrasonic tips was least for the Legacy and greater for the WhiteStar and Millennium under all conditions. The causes of these findings appear to be the longer stroke length of the WhiteStar and Millennium and the underestimation of the duty cycle with the WhiteStar.

Facebook səhifəmizə qoşulun

Elm tərəfindən dəstəklənən ən tam dərman bitkiləri bazası

  • 55 dildə işləyir
  • Elm tərəfindən dəstəklənən bitki mənşəli müalicələr
  • Təsvirə görə otların tanınması
  • İnteraktiv GPS xəritəsi - yerdəki otları etiketləyin (tezliklə)
  • Axtarışınızla əlaqəli elmi nəşrləri oxuyun
  • Təsirlərinə görə dərman bitkilərini axtarın
  • Maraqlarınızı təşkil edin və xəbər araşdırmaları, klinik sınaqlar və patentlər barədə məlumatlı olun

Bir simptom və ya bir xəstəlik yazın və kömək edə biləcək otlar haqqında oxuyun, bir ot yazın və istifadə olunan xəstəliklərə və simptomlara baxın.
* Bütün məlumatlar dərc olunmuş elmi araşdırmalara əsaslanır

Google Play badgeApp Store badge